Rational installation and use of vibration Testing Chambers
Jan 10, 2018

Rational installation and use of vibration Testing Chambers

1. The correct use of the vibration Testing Chambers:

1, the vibration Testing Chambers must comply with the requirements that the voltage distribution, the same specifications and the vibration Testing Chambers input table and a control box outside the box for the ground terminal, it must be properly grounded, to ensure safety.

2, first turn on the power switch, then the waveform for full wave, vibration direction (up and down, and finally adjust the parameters of vertical).

3, in order to make the power amplifier module and a cooling body sufficient time, must cut the signal after cooling can disconnect the power amplifier circuit breaker leakage was 7 to 10 minutes.

4, when the vibration Testing Chambers works, do not close the magnetic or unsuitable magnetic objects (such as watches, such as watches) near the vibration generator.

5, it is forbidden to close the control box and the microcomputer power before closing the amplifier, otherwise it will cause the impact or even damage to the power amplifier and the vibrating table.

Two. The reasonable installation of thevibration Testing Chambers:

1, environmental temperature and humidity conditions

Environmental temperature requirements: 5 to 35 degrees centigrade

Relative humidity: less than 85%RH

Environmental pressure requirements: 86KPa ~ 106KPa

Storage safety conditions: when the vibrating table is not working, the storage environment of the equipment should be kept more than 0.

2. Installation site and size requirements

Bearing weight: ground bearing requirements of more than 350kg/ square meters

Channel requirements: according to the size of the vibrating table in the technical specification, it can be passed.

The ground is flat and well ventilated. There is no inflammable and explosive corrosive gas and dust. There is no strong electric magnetic field and radiation source in the vicinity. There is a proper maintenance space around the equipment.