My test chamber's temperature pull up is above the controller set point
Jan 11, 2018

The control is satisfactory however is at incorrect temperature.

Check this possible cause and solution:

The temperature controller may be out of calibration.

→ Calibrate the temperature controller.

The heater is energized continuously. This may result in damage to the test chamber and its contents.

Check these possible causes and solutions

The temperature controller of the unit is defective.

→ Replace the defective controller.

The relay is defective. This occurs only with models that have relays.

→ Replace the defective relay.

The heater has shorted to ground.

→ Determine ground and correct.

As a “Rising star” in the field of environmental testing, Hardy not only obtained certifications such as ISO9001:2008, ISO14001:2004, CE etc., but also owned several rewards, honour and 17 Chinese Patents awarded by Chinese government such as innovation design, optimal control etc. 

As a “Main force” in the environmental testing domain,Hardy provides products to customers such as climate testing chamber、Temperature Humidity Testing Chamber 、Environmental Stress Screening (ESS) Chamber、AGREE chamber、Temperature shock chamber、Walk-In Environmental Test Chamber、Altitude (Hypobaric) Test Chamber etc.Besides the regular testing products, Hardy has experience in designing and producing of multi-function large testing chamber and successful cases such as multi-function large vehicle drive in chamber, explosion safe chamber with each level of protection and function used for altitude environmental simulation, soldier training. The product application of Hardy involves in many such as electron industrial, engineering, machining, vehicle, biology, pharmacy, aviation, aerospace, military and alternative energy etc.